Audit Logging

Purpose

Define what and how to log for medical devices: security/safety events, integrity, protection, and retention of logs for compliance and forensics.

When to Apply

  • Security-relevant actions, configuration changes, access to PHI, safety events.
  • System/device state changes, updates, connectivity, and error conditions.

Requirements (testable)

  1. Scope: Log authentication events, config changes, safety actions, updates, anomalies, and PHI access (without logging PHI content). Rationale: accountability.
  2. Integrity: Protect logs against tampering (hash chain/MAC, restricted access); no user-level editing. Rationale: trustworthiness.
  3. Time Sync: Ensure timestamps are reliable (NTP/PTP or hardware RTC); detect drift. Rationale: ordering and forensics.
  4. Minimal PHI: Do not store PHI unless necessary; if stored, encrypt and minimize. Rationale: privacy.
  5. Retention: Define retention and secure storage/backup; support export for investigations. Rationale: compliance and availability.
  6. Performance: Logging must not jeopardize safety/timing; use buffering with backpressure limits. Rationale: safety.

Recommended Practices

  • Use structured logs (key=value/JSON) with event IDs.
  • Separate audit log from operational logs; stricter controls on audit.
  • Rate-limit repetitive events to avoid flooding; but ensure critical events always logged.
  • Include device ID, software version, and correlation IDs.

Patterns

Hash-chained log entry:

c
// REQ-AUD-INT-01; TEST-AUD-04
typedef struct {
    uint8_t prev_hash[32];
    audit_event_t ev;
    uint8_t hash[32];
} audit_record_t;

Structured log example:

text
evt=auth_fail user=clinician1 code=LOCKOUT ts=2026-01-04T12:00:00Z

Anti-Patterns (risks)

  • Logging PHI verbatim -> risk: privacy breach.
  • Editable or deletable logs -> risk: tampering and noncompliance.
  • Unsynchronized clocks -> risk: unusable forensic timeline.
  • Logging so much that timing is affected -> risk: safety impact.

Verification Checklist

  • Audit events defined and implemented (auth, config, safety, update, PHI access).
  • Log integrity protection implemented; access restricted.
  • Timestamps reliable; drift detection in place.
  • PHI minimized/omitted; if present, encrypted and justified.
  • Retention/backup defined; export supported securely.
  • Logging performance impact assessed; backpressure/limits applied.

Traceability

  • REQ-AUD-### to TEST-AUD-###; logs tied to device ID/version; align with Part 11/ISO 14971 controls.

References

  • 21 CFR Part 11 (audit trail), HIPAA Security Rule.
  • FDA cybersecurity guidance (logging/monitoring).
  • IEC 62443 logging recommendations (informative).

Changelog

  • 1.0.0 (2026-01-04): Initial audit logging skill with scope, integrity, and retention guidance.

Audit History

  • 2026-01-04: Audit performed. Verified:
    • 21 CFR Part 11 audit trail requirements correctly referenced
    • HIPAA Security Rule reference appropriate
    • IEC 62443 logging recommendations correctly marked as informative